"Specs at a glance
Max. measurement error
Volume flow (standard): ±0.5 % o.r.± 1 mm/s (0.04 in/s)
Volume flow (option): ±0.2 % o.r. ± 2 mm/s (0.08 in/s)
Measuring range
4 dm³/min to 9600 m³/h (1 gal/min to 44 000 gal/min)
Medium temperature range
Liner material PFA: –20 to +150 °C (–4 to +302 °F)
Liner material PTFE: –40 to +130 °C (–40 to +266 °F)
Max. process pressure
PN 40, Class 300, 20K
Wetted materials
Liner: PFA; PTFE
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022);
Tantalum; Platinum; Titanium" "Field of application
Promag P is dedicated to chemical and process applications with
corrosive liquids and high medium temperatures. With its
straightforward hard- and software design, Promag P 10 simplifies
every step in its life cycle from engineering to servicing at usual
Endress+Hauser quality. Heartbeat Technology enables safe
processes.
The bidirectional measuring principle is virtually independent of
pressure, density, temperature and viscosity.
Dedicated to chemical and process applications with corrosive
liquids.
Device properties
Nominal diameter: max. DN 600 (24"")
All common Ex approvals
Liner made of PTFE or PFA
System integration with HART, Modbus RS485" "Benefits
Diverse applications – wide variety of wetted materials
Energy-saving flow measurement – no pressure loss due to cross
section constriction
Maintenance-free – no moving parts
Optimum usability – operation with mobile devices and SmartBlue app
or display with touch screen
Simple, time-saving commissioning – guided parameterization in
advance and in the field
Integrated verification – Heartbeat Technology" "Measuring
principle
Electromagnetic
Product headline
Flowmeter for basic process applications with easy-to-use operation
concept.
Dedicated to chemical and process applications with corrosive
liquids.
Sensor features
Diverse applications – wide variety of wetted materials.
Energy-saving flow measurement – no pressure loss due to cross
section constriction. Maintenance-free – no moving parts.
Nominal diameter: max. DN 600 (24""). All common Ex approvals.
Liner made of PTFE or PFA.
Transmitter features
Optimum usability – operation with mobile devices and SmartBlue app
or display with touch screen. Simple, time-saving commissioning –
guided parameterization in advance and in the field. Integrated
verification – Heartbeat Technology.
System integration with HART, Modbus RS485. Flexible operation with
app and optional display.
Nominal diameter range
DN 15 to 600 (½ to 24"")
Wetted materials
Liner: PFA; PTFE
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022);
Tantalum; Platinum; Titanium
Measured variables
Volume flow, conductivity, mass flow
Max. measurement error
Volume flow (standard): ±0.5 % o.r.± 1 mm/s (0.04 in/s)
Volume flow (option): ±0.2 % o.r. ± 2 mm/s (0.08 in/s)
Measuring range
4 dm³/min to 9600 m³/h (1 gal/min to 44 000 gal/min)
Max. process pressure
PN 40, Class 300, 20K
Medium temperature range
Liner material PFA: –20 to +150 °C (–4 to +302 °F)
Liner material PTFE: –40 to +130 °C (–40 to +266 °F)
Ambient temperature range
Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)
Flange material stainless steel: –40 to +60 °C (–40 to +140 °F)
Sensor housing material
DN 15 to 300 (½ to 12""""): AlSi10Mg, coated
DN 350 to 600 (14 to 24""""): Carbon steel with protective varnish
Transmitter housing material
AlSi10Mg, coated
Degree of protection
Standard: IP66/67, Type 4X enclosure
Display/Operation
2.4"" LCD display with touch & auto rotate; Configuration and
operation via Bluetooth technology possible
Outputs
4-20 mA HART (active/passive), Pulse/frequency/switch output
Modbus RS485, 4-20 mA
Digital communication
HART, MODBUS RS485
Power supply
DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area)
Hazardous area approvals
ATEX, IECEx, cCSAus, EAC, NEPSI, INMETRO, IND Ex, KC, JPN, UK Ex
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to
ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement
traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV
SÜD attestation)
Pressure approvals and certificates
PED, CRN
Material certificates
3.1 material
Hygienic approvals and certificates
Driking water approvals: ACS; NSF61, WRAS"